The Spaulding Classification met a key unmet need that today still forms the basis of international medical device disinfection guidelines.
The Spaulding Classification stratifies the risk of infection transmission based on the patient tissue the device will contact during use. The device classification determines the level of disinfection/sterilisation required (Table 1).
Table 1: Overview of The Spaulding Classification.1-6
Spaulding Classification | Medical Device Contacts | Risk of Infection Transmission | Disinfection Level |
| Critical | Sterile tissue or the bloodstream | High | Sterilisation* |
Semi-critical | Mucous membranes or non-intact skin | Medium | High-Level Disinfection (HLD) |
Non-critical | Intact skin only | Low | Intermediate-level (ILD) or Low-level disinfection (LLD) |
* Critical ultrasound probes can be high level disinfected and used with a sterile sheath if sterilisation is not possible.5,6
Sterilisation destroys all microorganisms.4
Critical devices must be sterile when used.1


The Spaulding Classification stratifies the risk of infection transmission based on the patient tissue the device will contact during use. The device classification determines the level of disinfection/sterilisation required (Table 1).
HLD destroys all microorganisms with the exception of high numbers of bacterial spores.4 A high level disinfectant is therefore bactericidal, virucidal (both lipid and non-lipid viruses), fungicidal and mycobactericidal.
Semi-critical ultrasound probes must undergo HLD and be used with a sheath.4-5
Critical ultrasound probes that cannot be sterilised can also undergo HLD.4,5 They must also be used with a sterile sheath.4, 5
An intermediate level or low level instrument grade disinfectant may be used for disinfection of non-critical instruments.4
Low level instrument grade disinfectants kill vegetative bacteria, some fungi and some viruses.4
Intermediate level instrument grade disinfectants kill vegetative bacteria, mycobacteria, viruses and most fungi but do not kill bacterial spores.4

It is important to recognise there are differences between disinfection levels and sterilisation. Correctly applying The Spaulding Classification to medical devices is a key part of keeping patients safe from healthcare-associated infections (HAIs).